UV Preconditioning Tester
|
|

|
 |
|
|
|
|
Product Overview
This Tester complies with the IEC 61215, IEC 61646 UV Preconditioning Test Specification. The unit’s radiation intensity is linearly adjustable and the intensity deviation is guaranteed to be within ±15% on the testing surface. It is equipped with a timer control with a range of 0 – 9999 hours, as well as a time accumulating and power shutdown with the ability to save data. It uses the standard PT-100 membrane thermo-transmitter to accurately control temperature in testing process
Technical Specification
| Item | Specification |
| Effective radiation area |
250×150 cm (adjustable) |
| Radiation uniformity |
±15% (to specimen’s surface) |
| Temperature control precision |
±2.0% |
| Centre-to-centre distance between light tubes |
70mm |
| Distance away light |
50~200mm adjustable |
| Temperature resolution |
0.1°C |
| Light-exposed temperature range |
RT+5~80°C/allowance ±5°C |
| Environment |
5~35°C low RH/300mm away from wall |
| Temperature control |
PID SSR |
| Box inner size |
L 200 × D 400 × H 200 cm |
| Weight |
1200kg |
| Power demand |
AC 220V±5% 50/60Hz 30A |
Features
- Comply with IEC 61215, IEC 61646 UV Preconditioning Test Specification
- Radiation intensity is linear adjustable; the intensity deviation is guaranteed within ±15% on the testing surface
- Test cabinet is equipped with timer controller in a time range of 0~9999 hr, as well as time accumulating and power-shutdown data-saving ability
- Using the standard PT-100 membrane thermo-transmitter to accurately control temperature in testing process
Applications
- Automotive
- Ground Transportation
- Defence/Military
- Aerospace
- Test Labs
- Production Testing
- Quality & Assurance Testing
- Fatigue Testing